CellFit v1.4.3
Download
Operation

Test & Sample Data

After opening a file, complete the test data. Some values are imported from .cfl files; others must be entered.

Filtrate

Field Purpose
Mass density Converts filtrate mass to volume. Required for volume results.
Dynamic viscosity Used in the specific resistance calculation.
Warning

Without the filtrate mass density, volume and any volume-based result remain unavailable.

Cake

Field Purpose
Dry mass Used in the specific resistance denominator.
Wet mass Used for dryness and related cake properties.

Other cake characteristics — dryness, thickness, surface mass — are reported when the underlying inputs are present.

Test conditions (from .cfl)

For .cfl files these are imported automatically: operator, sample reference, fluid, filter, filtration area, pressure, temperature, and loaded mass.

Note

For .cfc files, the filtration cell area defaults to the standard test cell. Verify it matches your setup.